Compatible Vacuum Gauges and Controllers

PSG500 tungsten
Web icon

Description

The INFICON Pirani Standard Gauges, PSG500 and PSG502-S, PSG510-S and PSG512-S employ the most advanced digital Pirani technology available in the marketplace. The rugged stainless steel sensor cell and compact design qualifies them for use on semiconductor systems and for standard applications, such as fore vacuum lines. The PSG500 / PSG502-S Pirani Gauge series are direct replacements for TTR 91 S / TTR96 S series: Part-No Typ Part-No Typ 350-060 PSG500 DN 16 ISO-KF = 230 035 TTR 91 DN 16 ISO-KF 350-061 PSG500 1/8" NPT = 230 038 TTR 91 1/8" NPT 350-062 PSG500 DN 16 CF-R = 230 036 TTR 91 DN 16 CF-R 350-068 PSG500 DN 16 CF-R, long = 230 037 TTR 91 DN 16 CF-R, long 350-080 PSG500-S DN 16 ISO-KF = 230 040 TTR 91 S DN 16 ISO-KF 350-081 PSG500-S 1/8" NPT = 230 043 TTR 91 S 1/8" NPT 350-088 PSG500-S DN 16 CF-R, long = 230 042 TTR 91 S DN 16 CF-R, long 350-140 PSG502-S DN 16 ISO-KF = 230 045 TTR 96 S DN 16 ISO-KF 350-148 PSG502-S DN 16 CF-R, long = 230 047 TTR 96 S DN 16 CF-R,...

Product characteristics

Measuring principle
thermal conductance according to Pirani
1 x 10 -3 … 100 mbar
±15%
5 x 10 -4 … 1 x 10 -3 mbar
±50%
100 … 1000 mbar
±50%
1 x 10 -3 … 100 mbar
2%
Voltage range
0 … +10.3
Measurement range
+1.9 … +10.0
Measurement range (air, 0 2, CO, N 2)
5 x 10 -4 to 1000
Number of setpoints
none
At gauge
+14 …+30
Ripple
≤1
Voltage vs. pressure
logarithmic 1.286 Volts per decade
DN 16 ISO-KF, DN 16 CF-R, 7/16-20 UNF
1.5 (0.092)
DN 16 ISO-KF & DN 16 CF-R long tube
10 (0.61)
1/8'' NPT , 4 VCR, 8 VCR, 1/2'' tube
2 (0.122)
Operation
+5 … +60
Vacuum connection (1)
80 / 250 (2)
Storage
-20 … +65
DN 16 ISO-KF, 7/16-20 UNF
80
DN 16 CF-R, 4 VCR
100
1/8'' NPT, 1/2'' tube
70
8 VCR, DN 16 ISO-KF long tube
130
DN 16 CF-R long tube
140
Error signal
0 … +0.5 (filament rupture)
Output impedance
2 x 4.7
Minimum loaded impedance
10, short-circuit proof
Response time
80
Gauge identification
27.0, referenced to supply common
Adjustment
one tactile switch for ATM and HV adjustment
Power consumption
≤1
Electrical connection
FCC 68 / RJ45 appliance connector, 8 poles, male
Sensor cable
8 poles plus shielding
Cable length
≤100 (8 x 0.14 mm 2)
Materials exposed to vacuum
glass, Ni, NiFe, DIN 1.4301 / 1.4305 / 1.4435
Admissible pressure
10, limited to inert gases
Mounting orientation
any
Degree of protection
IP40

Documents

Domain icon Manufacturer/ Producer

50968 Köln - Germany

Contact